Automate Agilent GC

Automate Agilent GC Calibration

A GasMix dynamic diluter and an Agilent GC can run a full multi-point calibration on their own, with no operator between steps. The two instruments talk over a simple digital I/O handshake: the GasMix waits until the GC is READY, triggers the run, holds each concentration for a user-set mixing and sampling time, then moves to the next level. In a documented coupling with an Agilent 490 Micro GC, CO2 calibration over 5 to 300 ppm gave a linearity of R² = 0.9999, a detection limit at the lowest point of 5 ppm, and repeatability of 2.6 % RSD on concentration (n = 20).

R² 0.9999CO2 linearity, 5 to 300 ppm
5 ppmdetection limit (lowest calibrated point)
2.6 %concentration RSD at 50 ppm (n=20)
0.16 %retention-time RSD (n=20)

Multi-point GC calibration is repetitive, time-consuming and easy to get wrong by hand. Coupling a dynamic gas diluter to the GC turns it into an unattended sequence: one certified cylinder becomes a full ladder of concentrations, generated and injected automatically. This page explains how the coupling works, what a published Agilent application note actually measured, and how to set it up with current Agilent instruments.

See the diluter built for this → AIOLOS III

Why automate GC calibration

Calibrating a GC across several concentration levels usually means swapping cylinders, waiting for lines to stabilise, launching each run and transcribing results, over and over. Every manual step adds time and a chance for error. Automating it with a dynamic diluter changes the economics:

  • One cylinder, many pointsA single higher-concentration certified standard is diluted on the fly to every level you need, instead of a cylinder per concentration.
  • Unattended sequencesThe diluter and GC coordinate themselves, so a full multi-point curve runs with no operator between steps.
  • Repeatable timingUser-set mixing and sampling times mean every level is handled identically, run after run.
  • Traceable outputConcentrations come from calibrated mass flow control with documented uncertainty, not hand-mixing.

How the GasMix and Agilent GC coupling works

The coupling relies on a bidirectional digital I/O handshake between the diluter and the GC, plus a stream selector to switch between calibration gas and sample. In the documented Agilent 490 Micro GC setup, it is wired like this:

Certified cylinder GasMix dynamic diluter + dilution gas Stream selector Agilent GC Micro GC / GC + detector gas flow READY out START trigger (relay) digital I/O
Coupling of a GasMix diluter with an Agilent Micro GC: gas flows left to right; the diluter and GC coordinate over a READY / START digital I/O handshake.
1GenerateGasMix dilutes the calibration gas to the target concentration, with a user-set mixing time to flush the line and injector.
2Wait for READYGasMix holds until the GC signals READY, so a run never starts before the analyser is prepared.
3Trigger STARTA relay triggers the GC START, with a user-selectable sampling time.
4Switch streamA VICI stream selector switches between calibration and sample; the pump aspirates at ambient pressure.
5Next levelThe system steps to the next concentration and repeats, building the full curve automatically.

Because the handshake is a generic READY / START digital I/O exchange, it is not specific to one GC model. It works with Agilent instruments that expose ready and start signals, current Micro GC and GC models included, as well as other analysers. On the software side, the documented setup used bidirectional communication between the Micro GC (Agilent OpenLAB CDS) and the GasMix (its own control software), today handled by AlySoft.

Automate your Agilent GC calibration

Tell us your GC model, target compounds and concentration range, and we will confirm the coupling and the diluter configuration for your setup.

Request a quote Start the Product Finder

What the Agilent application note measured

The results below come from an Agilent application note, Multi-point Gas Calibration, Coupling of a GasMix with a 490 Micro GC (Remko, Application Specialist Micro GC, Agilent Technologies). CO2 was analysed on a COx micro-packed column with argon carrier and a micro thermal conductivity detector, using the GasMix to dilute the calibration gas across the range.

Linearity and detection limit

MetricResult
CompoundCarbon dioxide (CO2)
Calibration range5 to 300 ppm
LinearityR² = 0.9999
Detection limit5 ppm (lowest calibrated point)
Column1 m COx micro-packed, heated injector, backflush
Carrier gasArgon

Calibration points were generated automatically at 5, 10, 50, 100, 150, 200 and 300 ppm. The note observes that the micro-packed COx column causes some peak broadening, and that the GasMix was used to reach the system's limit of detection at the low end.

Repeatability at 50 ppm

Parameter (CO2, 50 ppm, n = 20)RSD
Retention time0.16 %
Concentration (external standard method)2.6 %
The application note's conclusion: easy and convenient coupling of the Micro GC and GasMix is possible, with bidirectional communication between the two, and excellent linearity together with good retention-time and concentration repeatability.

Full details are in the Agilent application note: GasMix coupling with the 490 Micro GC (PDF).

Manual vs. automated GC calibration

CriterionManual calibrationGasMix + Agilent GC
Concentration levelsOne cylinder per levelFull ladder from one certified cylinder
Operator involvementPresent at every stepUnattended once the sequence starts
Run triggeringManual start, timing variesREADY / START handshake, identical each run
Line stabilisationJudged by the operatorUser-set mixing and sampling times
Low-level pointsHard to source as cylindersGenerated by dilution, down to the LOD
TraceabilityDepends on manual recordsCalibrated MFC control, documented

Setting it up with your Agilent instrument

The coupling needs three things: a diluter with digital I/O, a stream selector, and the ready/start wiring to the GC. For automated multi-point work the natural choice is AIOLOS III, which drives multi-gas dilution and sequences through built-in digital I/O; for compact two-gas dilution, ZEPHYR II; and for complex mixtures up to 16+ channels, HURRICANE. The same approach underpins trace-level gas generation and the broader question of accurate, traceable results. Not sure which diluter fits your GC? Choosing the right gas mixer walks through it, or the Product Finder recommends a configuration.

Frequently asked questions

Which Agilent GCs can be synchronised with a GasMix?

The coupling uses a generic READY / START digital I/O handshake, so it is not tied to one model. It works with Agilent instruments that expose ready and start signals, including current Micro GC and GC models, and with other analysers. The documented reference setup used an Agilent 490 Micro GC with OpenLAB CDS.

How does the diluter know when to start the GC run?

The GC signals READY when it is prepared; the GasMix waits for that signal, then uses a relay to START the GC. Before triggering, the GasMix holds the target concentration for a user-selectable mixing time so the line and injector are homogeneously flushed. A sampling time is also user-selectable.

What performance was actually demonstrated?

In the Agilent application note, CO2 was calibrated from 5 to 300 ppm with a linearity of R² = 0.9999 and a detection limit of 5 ppm. Repeatability at 50 ppm (n = 20) was 0.16 % RSD on retention time and 2.6 % RSD on concentration by the external-standard method.

Do I still need certified calibration gas?

Yes. A certified cylinder remains your traceable reference. The diluter extends it: from one higher-concentration standard it generates every intermediate level on demand, so you calibrate across a full range without buying a separate cylinder for each concentration.

How many calibration points can be run automatically?

As many as your method needs. In the reference note, seven CO2 levels (5, 10, 50, 100, 150, 200 and 300 ppm) were generated automatically in sequence. The diluter steps through each level, waits for the GC, triggers the run and moves on, unattended.

Is the automated calibration traceable?

The concentrations are produced by calibrated mass flow control rather than manual mixing, with documented uncertainty, and the diluter software logs the sequence. Combined with a certified source cylinder, this gives a defensible, traceable multi-point calibration. See how modern gas mixers achieve accurate, traceable results.

Turn GC calibration into an unattended sequence

25+ years specifying gas dilution systems. Tell us your Agilent GC and target compounds, and we will spec the coupling, or start with the Product Finder.

Request a quote Start the Product Finder

Or email gasmix@alytech.fr.

Reference: Multi-point Gas Calibration, Coupling of a GasMix with a 490 Micro GC, Remko, Agilent Technologies. Figures cited are from that note. Agilent and 490 Micro GC are trademarks of Agilent Technologies.

Scroll to Top